Results per page: 10 | 25 | 50 18 Pytania found, displaying all Pytania .
# Pytanie Odpowiedź Data ostatniej publikacji Status
18
In message #17, JRC wrote: "Information about authorised persons and opening procedure: Persons authorised [...]
As indicated in point 7 of the document 'Invitation to Tender' the request is to be sent to the following [...]
07/12/2017
17
Please take note that the date and time of the public opening will remain as indicated in the Contract [...]
Conditions for opening of tenders Date: 14/12/2017 Local time: 14:00 Place: The meeting point for the [...]
07/12/2017
16
Can the deadline be extended?
We kindly ask you to see the information provided in question 13.
06/12/2017
15
From the technical specification of Lot 1 I read that IDAS system must be connected to IRDS defined in [...]
Please find the answers below: 5) The first answer is to be found in point 6.1.0.1 c: "assembly of all [...]
04/12/2017
14
From the technical specification of Lot 1 I read that IDAS system must be connected to IRDS defined in [...]
Please find below the answers to the questions above: 1) Each HpGe has 2 output signals: 1 for energy [...]
04/12/2017
13
Information about tender deadline and technical support.
The Tender deadline will be extended shortly. The exact date and time of the new deadline will be announced [...]
04/12/2017
12
We are inserting the documents on the TED system but the system does not accept the documents with extension [...]
A P7M file is an encrypted message and these type of document are not supported by the TED system. Therefore [...]
04/12/2017
11
If I want to submit the tender not as solo but with a group (joint tender, subcontractors or third party [...]
For each selection criterion, whether financial (section 2.3.2 of the administrative annex) or technical [...]
29/11/2017
10
Please note that the revision 2, dated 28/11/2017 of the technical specifications "Acquisition of radiometric [...]
Please note that the revision 2, dated 28/11/2017 of the technical specifications "Acquisition of radiometric [...]
28/11/2017
9
Please provide required length of BNC cables.
The minimum distance between the VETO detectors and the IDAS (Integrated Digital Analysis System – see [...]
28/11/2017
8
Is it expected that only plastic detectors with PMT Bases+BNC cables should be delivered and installation [...]
See § 6.1.0.1 c) of the technical specification: "The supply of the detection system consists of (c) [...]
28/11/2017
7
Question to LOT 7: In supply description (chapter 5) is stated „The size of the whole structure (instrument [...]
The full size (including any accessories or tools) of the instrument is fixed due to the space constraints [...]
22/11/2017
6
Can you please confirm that the size of the whole structure (75 x 58 x 76 cm +- 5%) is inclusive of cart [...]
Yes, we confirm that the dimensions of the instrument specified in the technical specifications includes [...]
16/11/2017
5
System we will propose is compliant with all tech. specifications ( many of them are superior than the [...]
Please note that the technical specifications "Acquisition of radiometric equipment - Supply and installation [...]
10/11/2017
4
What shall we interpret the following statement? "Each detector shall be connected to...PMT bases with [...]
Each module of the VETO System includes a plastic scintillator, PMT and PMT base and shall have one connector [...]
09/11/2017
3
please inform if the ESTIMATED TOTAL VALUE of each lot can be exceeded?
The Tender specifications: Part 1- Administrative Annex states in point 2.4 that tenders may be considered [...]
08/11/2017
2
Can you please clarify whether the energy range required for the HpGe detectors (see page 7/15 of the [...]
Even if the probability of detecting 3 keV gamma rays could be low, considering that the detector will [...]
07/11/2017
1
Can you please state the maximum expected density of the material to be characterised?
The real density of the single piece of the objects to be measured can be up to 20 g/cm3 while the density [...]
07/11/2017