Results per page: 10 | 25 | 50 19 Questions found, displaying all Questions .
# Question Answer Last publication date Status
19
Please confirm that this is a complete mailing address for the hand singed documents: CALL FOR TENDERS [...]
European Commission - Joint Research Centre Resources Directorate – Unit R.8 Finance and Procurement [...]
29/06/2020
18
We read your bid documents but couldn't find an electronic access to bid on. Please give to us for your [...]
On the page https://etendering.ted.europa.eu/cft/cft-display.html?cftId=6230, left side, it is the link [...]
22/06/2020
17
I don't know how to prove the signature. Is it possible to make a copy of my passport?
It is not required evidence to prove the signature. The main rules regarding the signature are foreseen [...]
22/06/2020
16
We can not find the template from Annex 2 of Tender Specifications There is a blank page). Is any other [...]
The template of the Declaration on Honour on exclusion and selection criteria is uploaded on the website [...]
18/06/2020
15
If the number of bus bars in the sample stage is not decided, the estimated price will vary greatly. [...]
The answer for the question 14 has been corrected. The tenderer is not required to propose the contracting [...]
18/06/2020
14
Does the stage need to include bus bar contacts, and if so for how many bus bars?
The submitted answer has been corrected. An appropriate contacting system for such dimensions is not [...]
18/06/2020
13
We would like to use hand-written signature for all the documents of this tender that need to be signed [...]
The hand-signed originals must be sent by letter (meaning dispatch date not receipt date) to the contracting [...]
10/06/2020
12
Dear Sir/Madam, In order to provide a suitable system to hold the solar cell, we will design our sample [...]
The JRC will measure several different solar cells on the sample stage not a single type. The sample [...]
10/06/2020
11
3. Could you please clarify point d1. of you specifications? "Protection for safe electrical and irradiance [...]
The supplier is responsible for supplying an instrument which is compliant with the Italian and EU electrical [...]
10/06/2020
10
Can you send your sample reference Cells before our final testing, if you possible? A) We would like [...]
No it is not possible for the JRC to send our internal reference cells to the successful tenderer before [...]
10/06/2020
9
What is the manufacturer and model of the spectroradiometer owned by JRC? A) Because we have different [...]
The JRC has several Spectroradiometers, for this acceptance test the JRC will use either the Instrument [...]
10/06/2020
8
2. What is the definition of the shortest wavelength and the longest wavelength? (b5) A) Is it all right [...]
The definition is the lowest and highest wavelength which still as some irradiance signal present. Is [...]
10/06/2020
7
1. Please advise to your requested specifications for sample stage. (b.6) A) Do you need Vacuuming system [...]
The sample stage as stated in point b.6 is intended for crystalline devices ( which typically are bare [...]
10/06/2020
6
dear Sirs, We are representing as ............. the manufacturer of these solar simulators who is based [...]
The Commission will sign the contract with the company that has submitted the tender and the payments [...]
09/06/2020
5
Dear sirs and or madams, Would it be acceptable to propose an LED based system as opposed to the Xe bulb [...]
As stated in the Technical Specification the proposed system shall: a.1 Steady-state light source shall [...]
08/06/2020
4
1. TESTING: after installation and training, the JRC will perform tests in order to verify the technical [...]
The JRC will use its own equipment for testing the functionality and performances of the supply delivered [...]
05/06/2020
3
2. In the "Technical Specifications" document you write: "The contractor must demonstrate to the Commission [...]
This point requires that what the supplier presented in the offer is in fact delivered to the JRC and [...]
05/06/2020
2
dear Sirs, For projects like this we usually request a advance payment. For example, 30% T/T in advance [...]
In an open procedure, the terms of the procurement cannot negotiated. The payment arrangements shall [...]
28/05/2020
1
We have an additional question with regard to part 2 of the technical specifications. Under 2.1. Supply [...]
The sample stage as stated in point b.6 is intended for crystalline devices ( which typically are bare [...]
25/05/2020