70
• Performance Envelope: Defined as the air speed parameters for which the cabinet must operate, NSF requires [...]
Based on the market research carried out for the preparation of the technical specifications, this item [...]
17/08/2021
Răspuns primit
69
Technical specifications describe that "Should have an approximate volume of ..... open orifice maximum [...]
The following requirement “Should have an approximate volume of 370 Lt, and above platform volume of [...]
17/08/2021
Răspuns primit
68
Technical specifications describe that "Should have an approximate volume of 370 Lt ....." The device [...]
The following requirement “Should have an approximate volume of 370 Lt, and above platform volume of [...]
17/08/2021
Răspuns primit
67
Technical specifications describe that "The ULT freezer must have four inner doors." The device we want [...]
The following requirement “The ULT freezer must have four inner doors” will be replaced by “The ULT freezer [...]
17/08/2021
Răspuns primit
66
Technical specifications describe that "The ULT freezer must have ergonomic 7'' full touch screen, self [...]
Based on the market research carried out for the preparation of the technical specifications, this item [...]
17/08/2021
Răspuns primit
65
Technical specifications describe that "The ULT freezer must be horizontal model with minimum 400 L capacity [...]
The following requirement “The ULT freezer must be horizontal model with minimum 400 L capacity” will [...]
17/08/2021
Răspuns primit
64
Technical specifications describe that "The ULT freezer must have removable inventory racks to store [...]
Based on the market research carried out for the preparation of the technical specifications, this item [...]
17/08/2021
Răspuns primit
63
Technical specifications describe that "Ultra-low temperature freezer (ULT freezer) must have a temperature [...]
The following requirement “Ultra-low temperature freezer (ULT freezer) must have a temperature range [...]
17/08/2021
Răspuns primit
62
Technical specifications describe that "Cross-contamination: 0.01% max." The technical specification [...]
Based on the market research carried out for the preparation of the technical specifications, this item [...]
17/08/2021
Răspuns primit
61
Technical specifications describe that ‘‘Injection-volume accuracy: 1% max." The technical specification [...]
Based on the market research carried out for the preparation of the technical specifications, this item [...]
17/08/2021
Răspuns primit