70
|
• Performance Envelope: Defined as the air speed parameters for which the cabinet must operate, NSF requires [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
69
|
Technical specifications describe that "Should have an approximate volume of ..... open orifice maximum [...]
|
The following requirement “Should have an approximate volume of 370 Lt, and above platform volume of [...]
|
17/08/2021
|
|
68
|
Technical specifications describe that "Should have an approximate volume of 370 Lt ....." The device [...]
|
The following requirement “Should have an approximate volume of 370 Lt, and above platform volume of [...]
|
17/08/2021
|
|
67
|
Technical specifications describe that "The ULT freezer must have four inner doors." The device we want [...]
|
The following requirement “The ULT freezer must have four inner doors” will be replaced by “The ULT freezer [...]
|
17/08/2021
|
|
66
|
Technical specifications describe that "The ULT freezer must have ergonomic 7'' full touch screen, self [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
65
|
Technical specifications describe that "The ULT freezer must be horizontal model with minimum 400 L capacity [...]
|
The following requirement “The ULT freezer must be horizontal model with minimum 400 L capacity” will [...]
|
17/08/2021
|
|
64
|
Technical specifications describe that "The ULT freezer must have removable inventory racks to store [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
63
|
Technical specifications describe that "Ultra-low temperature freezer (ULT freezer) must have a temperature [...]
|
The following requirement “Ultra-low temperature freezer (ULT freezer) must have a temperature range [...]
|
17/08/2021
|
|
62
|
Technical specifications describe that "Cross-contamination: 0.01% max." The technical specification [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
61
|
Technical specifications describe that ‘‘Injection-volume accuracy: 1% max." The technical specification [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
60
|
Should include an autosampler with the following features: Technical specifications describe that ‘‘Number [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
59
|
Technical specifications describe that ‘‘Should include an FCV valve unit." The technical specification [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
58
|
Technical specifications describe that ‘‘Should include an online degasser with at least 3 channels and [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
57
|
"Technical specifications describe that ‘‘Pulsation: 0.1 MPa" The technical specification refers to a [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
56
|
Should include a solvent delivery unit with the following features: Technical specifications describe [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
55
|
Should have control software with the following features: Technical specifications describe that "Auto-tuning [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
54
|
Technical specifications describe that "Should have a detection system: pulse counting" The technical [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
53
|
Technical specifications describe that "Should have as collision cell: quadrupole type collision cell [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
52
|
Technical specifications describe that "Should have DL maximum temperature: 300 °C" The technical specification [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
51
|
Technical specifications describe that ‘Should have cross talk: < 0.005 %.‘’ The technical specification [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
50
|
Technical specifications describe that ‘’Should have sensitivity: ESI positive, 1 pg reserpine, S/N & [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
49
|
Technical specifications describe that ‘’ Should have mass range: m/z 10 to 2,000 ‘’ The technical specification [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
48
|
1) Item 13 Autoclave. Can you please confirm chamber volume? 420 litres is very big and the max chamber [...]
|
1) Based on the market research carried out for the preparation of the technical specifications, this [...]
|
17/08/2021
|
|
47
|
Item 14 Spectrophotometer Once again the specs are totally locked out for a particular product: https [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
46
|
Item 5 .- Class 2 Microbial Safety Cabinet Since this is a EuropeAid tender, it is required the goods [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
45
|
In particular: Item 2 - ICP-OES (Inductively coupled plasma - optical emission spectroscopy) system The [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
44
|
Dear Sirs Reference is made to project "REFORM/2021/EA-OP/0005 “ We bring to your attention the fact [...]
|
See our replies to questions 28, 29, 30 and 31.
|
17/08/2021
|
|
43
|
There is no flow cytometer device on the market that meets the origin requirement. For this reason, this [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
42
|
Technical specifications describe that “Should be a table top type, with circulation, and should work [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
41
|
Technical specifications describe that “Time for sealing should not exceed 2 seconds.” For devices that [...]
|
The following requirement “Time for sealing should not exceed 2 seconds.”will be replaced by “Time for [...]
|
17/08/2021
|
|
40
|
Technical specifications describe that “Android system, 7-inch capacitive touch screen, intuitive APP [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
39
|
Technical specifications describe that “In addition, the structure should effectively solve the instability [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
38
|
Technical specifications describe that “The lifting motor structure should make the liquid column stretch [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
37
|
Technical specifications describe that “To added the Fluorescence detection function, which can be used [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
36
|
Technical specifications describe that “For samples with concentrations below 2ng/ul, fluorometric mode [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
35
|
Technical specifications describe that "The sterilizer shall have single door (hinge type) with radial [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
34
|
Technical specifications describe that “It must have LIMS connection ability.” Products manufactured [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
33
|
Technical specifications describe that “The patient results image should provide additional information [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
32
|
Technical specifications describe that “Device status, plate ID, examination process name, starting and [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
31
|
Technical specifications describe that “The system should have two 5-filter advanced optical units and [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
30
|
Technical specifications describe that "The centrifuge should have a max. speed at least 15,000 rpm." [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
29
|
Technical specifications describe that "The system should be able to store up to minimum 90 user-defined [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
28
|
Technical specifications describe that "Should have a 8 place fixed angle rotor for 15 ml tubes with [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
27
|
Technical specifications describe that "Should have ability to spin for a time of 1 – 99 min, continuous [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
26
|
Technical specifications describe that "Should have max. capacity of not less than 8 x 15 ml." For a [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
25
|
Technical specifications describe that "Should have a 24-place fixed angle rotor for 1.5/2.0 ml tubes [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
24
|
Technical specifications describe that "Should provide temperature control from -10oC or lower to +40oC [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
23
|
Technical specifications describe that "Should have ability to spin for a time of 1 sec - min: 59 sec [...]
|
Based on the market research carried out for the preparation of the technical specifications, this item [...]
|
17/08/2021
|
|
22
|
Technical specifications describe that "Tanks should have outer diameter about 80 cm, height about 150 [...]
|
The following requirement “Tanks should have outer diameter about 80 cm, height about 150 cm, and empty [...]
|
17/08/2021
|
|
21
|
We could not be sure if we are allowed to offer products originating from the following countries: 1) [...]
|
Article 9 of the Aid Regulation (EC 389/2006) specifies that all supplies and materials purchased under [...]
|
17/08/2021
|
|