1
|
NELLA LIBRERIA DI DOCUMENTI E' PRESENTE LA DOCUMENTAZIONE SOLO IN LINGUA INGLESE SI RICHIEDE SE POSSIBILE [...]
|
Sì, è possibile. Potete indicarci a che lotto siete interessati?
|
01/06/2018
|
|
2
|
Buongiorno, con riferimento alla procedura di gara in oggetto, si chiede la cortesia di poter ricevere [...]
|
Comunichiamo che la traduzione in italiano sarà disponibile nei prossimi giorni.
|
01/06/2018
|
|
5
|
SIAMO INTERESSATI AL LOTTO 1 E LOTTO 3
|
Da oggi 7 giungo 2018 la documentazione di gara è disponibile anche in italiano.
|
07/06/2018
|
|
3
|
Is the lot 1 a full package? (so the same lab has to do all the parameters?) Or one lab can submit a [...]
|
Lot 1 is a full package. As explained in § 1.2 of the Administrative Annex offers can be submitted by [...]
|
06/06/2018
|
|
4
|
Is it possible to submit all documentation in French language relative to the Tender ? Can you confirm [...]
|
We confirm that offers can be submitted in French, as in any other EU official language.
|
07/06/2018
|
|
7
|
We kindly ask you to inform us about the LOQ for LOT1 especially for heavy metalls but also for basic [...]
|
Please find here after the limit of quantification (LOQ) in mg/kg for each heavy metal: As 2.84 Cd 0.07 [...]
|
22/06/2018
|
|
6
|
In the Technical Specification, it is mentioned that all 26000 soil samples will be prepared and ready [...]
|
Individual soil samples are held in labelled plastic bags that are stored in groups in cardboard boxes [...]
|
22/06/2018
|
|
12
|
You mentioned some LOQ for heavy metals that are exceptionally low for some compounds. Are they mandatory [...]
|
For the indicated substances, we can accept LOQ in the following ranges: Mercury: 0.00005 - 0.025 mg/ [...]
|
22/06/2018
|
|
13
|
In respect of the 24-month guarantee period applying to Lots 2 and 3; can the customer confirm its expectation [...]
|
A guarantee of two years is required on the accuracy of information if seemingly anomalous values become [...]
|
22/06/2018
|
|
10
|
Lot 1 - Heavy metals specifies that the solutions resulting form aqua regia digestion should be analysed [...]
|
As indicated in the Technical Specifications, the use of an ICP-MS would be acceptable provided that [...]
|
22/06/2018
|
|