12
|
You mentioned some LOQ for heavy metals that are exceptionally low for some compounds. Are they mandatory [...]
|
For the indicated substances, we can accept LOQ in the following ranges: Mercury: 0.00005 - 0.025 mg/ [...]
|
22/06/2018
|
|
9
|
What is expected time from opening offers, when contract will be signed and project started?
|
After the opening of offers, the evaluation process will take place. This full process up to contract [...]
|
22/06/2018
|
|
8
|
We kindly ask you to provide a full set of Tender documentation in French for the whole 3 lots. Is to [...]
|
The French tender documents are online from today, 20 June 2018.
|
20/06/2018
|
|
7
|
We kindly ask you to inform us about the LOQ for LOT1 especially for heavy metalls but also for basic [...]
|
Please find here after the limit of quantification (LOQ) in mg/kg for each heavy metal: As 2.84 Cd 0.07 [...]
|
22/06/2018
|
|
20
|
Since the proposed requirements for quantification of antibiotic resistance genes are extremely specific [...]
|
As indicated in the Technical Specifications, the use of alternative methods would be acceptable provided [...]
|
22/06/2018
|
|
5
|
SIAMO INTERESSATI AL LOTTO 1 E LOTTO 3
|
Da oggi 7 giungo 2018 la documentazione di gara è disponibile anche in italiano.
|
07/06/2018
|
|
11
|
Regarding Criterion T1:The tenderer must prove experience in the field of analysing plant protection [...]
|
The tenderer has to prove analysis of each group of PPPs and antibiotics in Table 2 of the Technical [...]
|
22/06/2018
|
|
15
|
On the 6th page there is the sentence: "For brand names indicated in the Technical Specifications, fully [...]
|
Brands refer to trademarks, standards or producers names.
|
22/06/2018
|
|
1
|
NELLA LIBRERIA DI DOCUMENTI E' PRESENTE LA DOCUMENTAZIONE SOLO IN LINGUA INGLESE SI RICHIEDE SE POSSIBILE [...]
|
Sì, è possibile. Potete indicarci a che lotto siete interessati?
|
01/06/2018
|
|
10
|
Lot 1 - Heavy metals specifies that the solutions resulting form aqua regia digestion should be analysed [...]
|
As indicated in the Technical Specifications, the use of an ICP-MS would be acceptable provided that [...]
|
22/06/2018
|
|