Gegevens van de oproep tot inschrijving

Titel:
Levering van een röntgendiffractometer.
Aanbestedende dienst:
European Commission, Joint Research Centre (JRC)
Publicatiedatum op TED:
01/09/2017
Termijn voor de ontvangst van inschrijvingen:
04/10/2017
Status:
Afgesloten
Filterparameters tonen

Results per page: 10 | 25 | 50 4 Vragen found, displaying all Vragen .
# Vraag Antwoord Laatste publicatie Status
4
Would a goniometer with fixed 600mm measuring diameter be treated equally?
Dear Sir/Madam, Yes, provided that the required flexibility regarding future potential implementing of [...]
22/09/2017
3
Should the focusing or parallel beam optics already be quoted as an option within the quote or should [...]
Dear Sir/Madam, The device shall be capable of implementing focusing or parallel beam optics at a later [...]
20/09/2017
2
"Radiation sensitive parts of the device which cannot be separated from the goniometer must be well shielded [...]
Dear Sir/Madam, the radiation sensitive parts which cannot be separated from the goniometer must be designated [...]
20/09/2017
1
Should the channel detector already be quoted as an option within the quote or should the system only [...]
Dear Sir/Madam, the channel detector must be quoted as an option and also clearly specified as an option [...]
20/09/2017