Detalii licitaţie

Titlu:
Supply of Goods to Support Diagnostic and Analytical Capacities of Health and Fo...
Autoritate contractantă:
European Commission, DG for Structural Reform Support (REFORM)
Data de publicare pe TED:
28/06/2021
Termen pentru primirea ofertelor:
30/08/2021
Status:
Închisă
Afişare parametrii de filtrare

Results per page: 10 | 25 | 50 70 Întrebări found, displaying 1 to 10.[First/Prev] 1, 2, 3, 4, 5, 6, 7 [Next/Last]
# Întrebare Răspuns Data ultimei publicări Status
70
• Performance Envelope: Defined as the air speed parameters for which the cabinet must operate, NSF requires[...]
Based on the market research carried out for the preparation of the technical specifications, this item[...]
17/08/2021
69
Technical specifications describe that "Should have an approximate volume of ..... open orifice maximum[...]
The following requirement “Should have an approximate volume of 370 Lt, and above platform volume of [...]
17/08/2021
68
Technical specifications describe that "Should have an approximate volume of 370 Lt ....." The device[...]
The following requirement “Should have an approximate volume of 370 Lt, and above platform volume of [...]
17/08/2021
67
Technical specifications describe that "The ULT freezer must have four inner doors." The device we want[...]
The following requirement “The ULT freezer must have four inner doors” will be replaced by “The ULT freezer[...]
17/08/2021
66
Technical specifications describe that "The ULT freezer must have ergonomic 7'' full touch screen, self[...]
Based on the market research carried out for the preparation of the technical specifications, this item[...]
17/08/2021
65
Technical specifications describe that "The ULT freezer must be horizontal model with minimum 400 L capacity[...]
The following requirement “The ULT freezer must be horizontal model with minimum 400 L capacity” will[...]
17/08/2021
64
Technical specifications describe that "The ULT freezer must have removable inventory racks to store [...]
Based on the market research carried out for the preparation of the technical specifications, this item[...]
17/08/2021
63
Technical specifications describe that "Ultra-low temperature freezer (ULT freezer) must have a temperature[...]
The following requirement “Ultra-low temperature freezer (ULT freezer) must have a temperature range [...]
17/08/2021
62
Technical specifications describe that "Cross-contamination: 0.01% max." The technical specification [...]
Based on the market research carried out for the preparation of the technical specifications, this item[...]
17/08/2021
61
Technical specifications describe that ‘‘Injection-volume accuracy: 1% max." The technical specification[...]
Based on the market research carried out for the preparation of the technical specifications, this item[...]
17/08/2021