1
|
We are interested with this tender and we would like to ask you to add into the Document Library, the [...]
|
The requested documents are now published in Word on eTED.
|
23/07/2021
|
|
2
|
We have contacted several manufacturers of such systems and they all inform us that it is very difficult [...]
|
The drawings/plans of the hospital with the position of the stations marked on them is now published [...]
|
23/07/2021
|
|
3
|
Dear Sirs, please inform us about the final site(s) of delivery for the equipment, is the final delivery [...]
|
As described in the Instructions to tenderers point 1.1: “1.1. The subject of the contract is the supply [...]
|
23/07/2021
|
|
4
|
Dear Sirs, We are unable to open the drawings pdf document. The document seems to be corrupted. Could [...]
|
Please, try now
|
23/07/2021
|
|
5
|
We have analyzed the documentation sent, but it needs an integration to verify the connection distances [...]
|
The drawings/plans and some pictures of the hospital with the position of the stations marked on them [...]
|
05/08/2021
|
|
6
|
It has been requested – “support stand with levelling screws adjustable from 55- 85 mm.” Will you agree [...]
|
According to PRAG, section 4.3.4 - Additional information during the procedure - “No prior opinion on [...]
|
05/08/2021
|
|
7
|
It has been requested “Magnehelic gauge are required to indicate pressure drop across HEPA filter” Will [...]
|
According to PRAG, section 4.3.4 - Additional information during the procedure - “No prior opinion on [...]
|
05/08/2021
|
|
8
|
It has been requested “Average air flow should be 90±20 fpm (down flow). Will you agree to accept Airflow [...]
|
According to PRAG, section 4.3.4 - Additional information during the procedure - “No prior opinion on [...]
|
05/08/2021
|
|
9
|
It has been requested “Must have dual-DC motors” Will you agree to accept single DC-ECM blower for 70% [...]
|
According to PRAG, section 4.3.4 - Additional information during the procedure - “No prior opinion on [...]
|
05/08/2021
|
|
10
|
It has been requested “Overall size of the cabinet should be approximately 1000 or 1300 x 825 x 2450 [...]
|
According to PRAG, section 4.3.4 - Additional information during the procedure - “No prior opinion on [...]
|
05/08/2021
|
|