9
|
What is expected time from opening offers, when contract will be signed and project started?
|
After the opening of offers, the evaluation process will take place. This full process up to contract[...]
|
22/06/2018
|
|
10
|
Lot 1 - Heavy metals specifies that the solutions resulting form aqua regia digestion should be analysed[...]
|
As indicated in the Technical Specifications, the use of an ICP-MS would be acceptable provided that [...]
|
22/06/2018
|
|
11
|
Regarding Criterion T1:The tenderer must prove experience in the field of analysing plant protection [...]
|
The tenderer has to prove analysis of each group of PPPs and antibiotics in Table 2 of the Technical [...]
|
22/06/2018
|
|
12
|
You mentioned some LOQ for heavy metals that are exceptionally low for some compounds. Are they mandatory[...]
|
For the indicated substances, we can accept LOQ in the following ranges: Mercury: 0.00005 - 0.025 mg/[...]
|
22/06/2018
|
|
13
|
In respect of the 24-month guarantee period applying to Lots 2 and 3; can the customer confirm its expectation[...]
|
A guarantee of two years is required on the accuracy of information if seemingly anomalous values become[...]
|
22/06/2018
|
|
14
|
How compliance with the applicable environmental, social and labour law obligations established by Union[...]
|
At this stage tenderers do not need to prove this compliance in their offer. The Contracting Authority[...]
|
22/06/2018
|
|
15
|
On the 6th page there is the sentence: "For brand names indicated in the Technical Specifications, fully[...]
|
Brands refer to trademarks, standards or producers names.
|
22/06/2018
|
|
20
|
Étant donné que les exigences proposées pour la quantification des gènes de résistance aux antibiotiques[...]
|
As indicated in the Technical Specifications, the use of alternative methods would be acceptable provided[...]
|
22/06/2018
|
|
18
|
Les participants à l'appel d'offres peuvent-ils offrir des amorces améliorées pour séquencer la région[...]
|
The use of an alternative high-throughput sequencing platform is not allowed. Tenderer shall use a single[...]
|
22/06/2018
|
|
19
|
Dans l'offre, la ligature des index Illumina est demandée pour chaque échantillon amplifié par PCR séparément[...]
|
As indicated in the Technical Specifications, the use of alternative methods would be acceptable provided[...]
|
22/06/2018
|
|